Capnographic monitoring of respiratory activity improves safety of sedation for endoscopic cholangiopancreatography and ultrasonography

Gastroenterology. 2009 May;136(5):1568-76; quiz 1819-20. doi: 10.1053/j.gastro.2009.02.004.

Abstract

Background & aims: The Joint Commission on the Accreditation of Healthcare Organizations recommends ventilation monitoring during procedural sedation for gastrointestinal endoscopy. We sought to determine whether intervention, based on a microstream capnography-based ventilation monitoring system that has been shown to function as an early warning system for hypoxemia, would decrease hypoxemia during endoscopy.

Methods: Subjects undergoing elective endoscopic retrograde cholangiopancreatography (ERCP) and endoscopic ultrasonography (EUS) under procedural sedation with a combination of opioid and benzodiazepine were randomly assigned to either a study arm in which the endoscopy team was blinded to capnography or an open arm in which the endoscopy team was prompted of capnographic changes. The primary end point was the occurrence of hypoxemia; secondary end points were the occurrences of severe hypoxemia, apnea, and oxygen supplementation.

Results: A total of 263 subjects were enrolled; 247 were analyzed for efficacy. The numbers of hypoxemic events in the blinded and open arms were 132 and 69, respectively (P < .001). Thirty-five percent of all hypoxemic events occurred with completely normal ventilation. Hypoxemia developed in 69% of patients in the blinded arm compared with 46% in the open arm (P < .001). Severe hypoxemia percentages in the blinded and open arms were 31% and 15% (P = .004), for apnea were 63% and 41% (P < .001), for oxygen supplementation were 67% and 52% (P = .02), and for recurrent hypoxemia after oxygen supplementation were 38% and 18% (P = .01), respectively.

Conclusions: Capnographic monitoring of respiratory activity improves patient safety during procedural sedation for elective ERCP/EUS by reducing the frequency of hypoxemia, severe hypoxemia, and apnea.

Trial registration: ClinicalTrials.gov NCT00675415.

Publication types

  • Randomized Controlled Trial

MeSH terms

  • Capnography*
  • Cholangiopancreatography, Endoscopic Retrograde*
  • Conscious Sedation* / adverse effects
  • Double-Blind Method
  • Endosonography*
  • Female
  • Humans
  • Hypoxia / diagnosis
  • Hypoxia / etiology
  • Male
  • Middle Aged
  • Monitoring, Physiologic*

Associated data

  • ClinicalTrials.gov/NCT00675415